We talk to a lot of customers, and hear many of the same questions. Here is a list of several of the most common, along with a brief answer. In-depth answers can be found in Caron’s Service manuals, Preventative Maintenance Kit instructions or Field Installed Accessory instructions.

Questions? Explore Caron's Scientific for life sciences and industry. Request a quote or call our expert team now.

All Caron chambers require pH-neutral water of between 50K Ohms and 1Meg Ohms resistivity. This type of water can be obtained from distilled, reverse osmosis (RO) or deionized (DI) sources. Caron offers a system, the CRSY condensate recirculator, for facilities without a house water system.

Less pure water, with resistivity less than 50K Ohms, can lead to difficult to remove mineral deposits on system components, or within the chamber. Water of greater purity, such as ultrapure or Water For Injection (WFI) above 1 Meg Ohm can attack key unit components, causing damage not covered under warranty.